26
OMNISCANNER
1. To set the reference, connect OMNIScanner and OMNIRemote using the test
cable and adapter supplied with the scanner (as shown).
2. Turn OMNIScanner on.
3. Press (Setup).
4. Select Set Reference.
5. On finding OMNIRemote, the scanner reports: Acquiring Reference Data. A
gauge will indicate the progress.
6. Once the reference values has been recorded, OMNIScanner displays the
REMOTE INFO screen.
7. Press ESC to return to OMNIScanner’s SETUP screen.
2.5 Project Selection
A project name must be selected for test-result management. Autotests and Cables
were previously assigned to a specific project through the use of the OMNIScanner
Configuration utility. (See Chapter 9, Uploading and Printing, for further
information about OMNIScanner Configuration.)
1. Use the up-arrow or down-arrow to highlight Project and press ENTER.
2. Select the appropriate Project from the drop-down list and press ENTER.